Benchtop ip-maldi-ms assay

Intervention

1 paper2 findings

  • 1 lab in 1 country
  • 2 of 2 findings with company ties

Outcomes

2

Assay trueness

upin humans1

1 study
  1. Benchtop ip-maldi-ms assay improves assay trueness in humans (100±20%).

    “Trueness of 100±20% was confirmed over the range from 15 to 150 pM for Aβ1‐40, and from 1.5 to 15 pM for Aβ1‐42 and APP669‐711”

    Humans

    Improved analytical performance of plasma amyloid β assay using immunoprecipitation‐mass spectrometry

    AAIC 2025 abstract24 Dec 2025

Assay reproducibility

upin humans1

1 study
  1. Benchtop ip-maldi-ms assay improves assay reproducibility in humans (CV < 5%).

    “Each biomarker values showed a coefficient of variation (CV) of less than 5% for both intra‐assay and inter‐assay reproducibility”

    Humans

    Improved analytical performance of plasma amyloid β assay using immunoprecipitation‐mass spectrometry

    AAIC 2025 abstract24 Dec 2025

Latest

Nov 2025: 0Dec 2025: 1Jan 2026: 0Feb 2026: 0Mar 2026: 0Apr 2026: 0May 2026: 0Jun 2026: 0Jul 2026: 0Aug 2026: 0Sep 2026: 0Oct 2026: 01 in 12 months