Sensors · Nov 26, 2020

Development and Validation of Open-Source Activity Intensity Count and Activity Intensity Classification Algorithms from Raw Acceleration Signals of Wearable Sensors

By Poitras I, Clouâtre J, Bouyer LJ et al.

BACKGROUND: A popular outcome in rehabilitation studies is the activity intensity count, which is typically measured from commercially available accelerometers. However, the algorithms are not openly available, which impairs long-term follow-ups and restricts the potential to adapt the algorithms for pathological populations. The objectives of this research are to design and validate open-source algorithms for activity intensity quantification and classification. METHODS: Two versions of a quantification algorithm are proposed (fixed [FB] and modifiable bandwidth [MB]) along with two versions of a classification algorithm (discrete [DM] vs. continuous methods [CM]). The results of these algorithms were compared to those of a commercial activity intensity count solution (ActiLife) with datasets from four activities (n = 24 participants). RESULTS: The FB and MB algorithms gave similar results as ActiLife (r > 0.96). The DM algorithm is similar to a ActiLife (r ≥ 0.99). The CM algorithm differs (r ≥ 0.89) but is more precise. CONCLUSION: The combination of the FB algorithm with the DM results is a solution close to that of ActiLife. However, the MB version remains valid while being more adaptable, and the CM is more precise. This paper proposes an open-source alternative for rehabilitation that is compatible with several wearable devices and not dependent on manufacturer commercial decisions.

doi.org/10.3390/s20236767Open accessPubMed 33256127AI & data

AgingResearch.News