Use of bioelectrical impedance to estimate skeletal muscle mass in older men
By de Lima LRA, Rech CR, Petróski EL
The objectives of this study are: a) to assess agreement between bioelectrical impedance (BIA) and dual-energy X-ray absorptiometry (DEXA) methods for estimating skeletal muscle mass (MME); b) to analyze the predictive power of anthropometric and BIA variables for predicting MME in older adults. A total of 60 older men (61 to 80 years), residing in southern Brazil, were evaluated. Anthropometric variables (body mass and height) were measured, resistance and fat-free tissue hydration variables were measured using the tetrapolar BIA technique (Biodinamics - BF-310), and a whole-body scan was also performed using DEXA (LUNAR PRODIGY DF + 14319 Radiation and software 7.52.002 DPX-L). The difference between the methods was assessed using the “t” test, residual analysis, and the correlation coefficient. The predictive value of anthropometric and BIA variables was assessed using multiple linear regression. BIA was observed to overestimate MME by an average of 0.6 kg (dp= 1.59) compared with DEXA; however, there was no statistically significant difference (p<0.05). A strong relationship between the methods was observed (r=0.90; p<0.01). Regression analysis demonstrated that the variable EST2/R explains 86% of the variation in MME when adjusted for body mass and age, and this relationship is independent of the variables body fat, fat-free tissue hydration, and IMC. Thus, the BIA method tested here is seen to be valid for estimating MME in older men, and its values can be better predicted by the proposed regression model based on the measurement of EST2/R adjusted for body mass and age.